Product Overview

Overview

The Yokogawa AQ6370D is a benchtop optical spectrum analyzer for characterizing telecom transmitters, laser spectra, and optical components from 600 to 1700 nm. Standard and High Performance versions offer a 0.02 nm minimum resolution setting. The High Performance version provides tighter wavelength accuracy from 1520 to 1580 nm.

Check the full model suffix and the adapters fitted to the optical input and reference output. The -L1 version has a built-in wavelength reference source for alignment and wavelength calibration. The -L0 version uses an appropriate external source for those operations.

Configuration options

Option Configuration What to check
-10 / -12 Standard ±0.02 nm wavelength accuracy from 1520 to 1620 nm under the specified test conditions.
-20 / -22 High Performance ±0.01 nm from 1520 to 1580 nm and ±0.02 nm from 1580 to 1620 nm under the specified conditions.
-01 / -02 Limited 600 to 1700 nm with a 0.05 nm minimum resolution setting. See the Limited Model supplement for accuracy and dynamic range.
-L0 / -L1 Reference source -L0 comes without an internal source. -L1 includes a wavelength reference. These suffixes are separate from the performance grade.

Confirm the AQ6370D grade and optical adapters

Request current availability, the offered unit’s full model suffix, photos of its optical ports and adapters, and available optical test records. Include your fiber and connector type, target wavelengths, and calibration certificate requirements.

Buying FAQ

Model FAQ

How do I tell a Standard AQ6370D from a High Performance unit?

Check the full code on the rear nameplate. Yokogawa identifies -10 and -12 as Standard, -20 and -22 as High Performance, and -01 and -02 as Limited. A Limited unit has a 0.05 nm minimum resolution setting, compared with 0.02 nm on the other two grades.

Will my FC or SC fiber connect to the AQ6370D?

The optical input takes an AQ9447 adapter in FC or SC. Check the installed input adapter and the polish of your connector. For the reference output on an -L1 unit, confirm the separate AQ9441 output adapter.

Does the built-in reference source replace a traceable calibration certificate?

The -L1 source is used for the instrument’s optical alignment and wavelength calibration. For an acceptance or audit record, request a separate calibration certificate for the offered serial number and confirm the calibration scope and test data you need.

Manufacturer lifecycle

Yokogawa discontinued the AQ6370D on December 1, 2023 and identifies the AQ6370E as its successor. Check the instruments’ performance and compatibility for your application.

Manufacturer documents

Technical Information

Specifications

Key Specifications

Wavelength range
600 to 1700 nm for the Standard, High Performance and Limited models.
Wavelength resolution settings
0.02 to 2.0 nm in seven settings on Standard and High Performance. Limited models have a 0.05 nm minimum setting.
Wavelength accuracy, 1520–1580 nm
±0.02 nm on Standard (-10/-12) or ±0.01 nm on High Performance (-20/-22) under Yokogawa’s specified fiber, warm-up, optical alignment, and wavelength calibration conditions.
Close-in dynamic range example
At ±1.0 nm from a 1523 nm peak and a 0.05 nm resolution setting, Yokogawa specifies 73 dB for Standard and High Performance and lists 78 dB as typical for High Performance only. Applies in the specified high-dynamic mode with resolution correction off.
Optical input
Supports single-mode and multimode fiber with a compatible input adapter. AQ9447 adapters are offered in FC or SC. Confirm the adapter and connector polish for the offered unit.
Wavelength reference
-L1 includes the internal reference source. -L0 uses an appropriate external reference for optical alignment and wavelength calibration. An optional AQ9441 adapter fits the -L1 source output.
Remote control
GP-IB, RS-232C and Ethernet (TCP/IP). Confirm compatibility with your control system.

Sales and Service

Warranty
6-Month Warranty
Calibration Status
Calibration certificate or test report available (as stated in your quotation)
Manufacturer-Listed Standard Accessories
Original manufacturer list: one regional power cord (none with suffix -Y); two A9088ZM rubber feet supplied as one sheet; one B8079YA manual CD; the Getting Started Guide, regional documents and worldwide contact list. The IM AQ6370D-51EN supplement applies only to the limited model. Manufacturer-listed standard accessories are a reference to the original model documentation. They do not guarantee the contents of a used instrument shipment. Confirm the actual supplied items for the specific unit before ordering.
Downloads

Documents

Why work with us?

We check the details with you

From specifications to accessories, we help you understand what's included.

Know what you're buying

We share what we know about the condition, along with any available records.

Tell us what you need

Have a specific requirement or deadline? We'll focus on what matters to you.

No need to start over

We keep track of your questions and previous discussions, so it's easy to pick up where we left off.

A closer look at AQ6370D

A closer look at AQ6370D

Open the product inquiry form